Tracheal tubes are intended to be inserted through the larynx into the trachea and tracheostomy tubes are designed for insertion into the trachea through a tracheostomy. These medical devices provide a patent airway in patients and maintain the air delivery from the mechanical ventilator to lungs. Tracheal tubes and tracheostomy tubes may be equipped with a cuff in order to secure the patient’s airway and prevent secretions from entering into lower respiratory tract. Both the designated size of the product is incompatible with the labelling and the cuff have a poor quality may affect the efficacy of mechanical ventilation or cause secondary injury to mucosa of the airway. The study established testing methods for tracheal tubes and tracheostomy tubes based on ISO 5361 and ISO 5366, including measurement of the designated size (inner diameter and outer diameter) and cuff resting diameter, along with cuff leak test, cuff herniation, cuffed tube collapse of tracheal tube and the security of 15 mm connector and neck-plate test of tracheostomy tube. The methods were also applied to 17 samples (4 tracheal tubes and 13 tracheostomy tubes). The results showed all the samples met the tolerance or specifications set by ISO standards except one tracheal tube’s marked value of cuff resting diameter is different from the registration. The violated product case was sent to administrative authorities for penalty. The results will be offered to competent authorities as regulatory references. |