英文摘要 |
This article describes the design and analysis of a deformable spinal cage, its coating with titanium nitride (TiN), and its biocompatibility testing. We designed the deforma ble spinal cage through an irreversible mechanism and explored its mechanical behavior through ANSYS simulation analysis. Additionally, we used vacuum coating technology to coat titanium alloys with TiN, thus enhancing the biocompatibility of 3D printed ti tanium alloys, which can create a porous surface that can be used in spinal fusion devices. The safety of the deformable vertebral fusion device was verified through ANSYS mechanical analysis. ANSYS simulation analysis revealed that the deformable spinal c age could bear a maximum stress of 110.86 MPa, which was lower than its failure strength of 254.78 MPa, demonstrating the structure’s safety. The accuracy of the ANSYS analysis was confirmed by comparing the simulated values with the experimental values fo r the spinal cage entity in a compression test. |