英文摘要 |
Intravascular catheters are intended to be transcutaneously inserted into the vascular system. These medical devices, including over-needle peripheral catheter and central venous catheter (CVC), are designed for infusion of drugs and nutrients, blood sampling or hemodynamic monitoring. Both are classified as“J.5200 Intravascular catheter”in the annex of Regulations Governing the Classification of Medical Devices. Recently, there are numerous adverse events concerning these medical devices, causing additional clinical treatments, most of which were caused by product leakage. This study aimed to establish testing methods for over-needle peripheral catheters and CVCs based on ISO 10555, ISO 11070 and ISO 9626, which included tests of corrosion resistance, peak tensile force, liquid leakage under pressure, air leakage into hub assembly during aspiration, strength of union of needle hub and needle tube, along with stiffness of tubing. These methods were also applied to 6 samples, inclusive of 3 over-needle peripheral catheters and 3 CVCs. The results showed that all the samples met the tolerance or specifications set by ISO standards, and will be offered to the competent authorities as regulatory references. |