英文摘要 |
The purpose of this study was to investigate the relevance between the invasivearterial pressure measurement value and noninvasive blood pressure measurementvalue. A total of 1,186 blood pressure values from 144 patients were obtained. Thestatistical procedures applied in data management (by Excel) were descriptive statisticsfrom EPI-INFO (6.01 edition). Repeat measurement analysis used GeneralizedEstimating Equations(GEE). The result indicated that there was significant relevancebetween these two kinds of blood pressure measuring methods. It indicated that thesystolic, diastolic and mean blood pressure value of noninvasive blood pressurewere equal to the systolic, diastolic and mean blood pressure value of invasive arterialpressure's 62 percent+ 31.6; 78.3 percent + 0.72; 70.6 percent + 8.8 before the monitorof invasive arterial pressure returned to zero. We therefore suggest that patientswho already set the invasive arterial pressure line do not need to measure two kindsof blood pressure at the same time. |