英文摘要 |
TDxR, EmitR d.a.u.(TM) and AbuSign(TM) commerically available kits for methamphetamine and opiate assays were evaluated for the effects of pH of urine and the addition of alum. Assays were tested at pH 2.0, 4.0, 6.0, 8.0, 10.0, 12 .0 and 1.0, 2.5, 5.0, 7.5% of alum. in urine. Results indicate that pH changes of urine have no effect on TDxR. False negative results ocurred at pH out of range 6.0-8.0 for AbuSign(TM) and pH lower than 6.0 for EmitR d.a.u.(TM) when methamphetamine concentration is 600 ng/ml. At pH 2.0 and methamphetamine at a concentration of 600, 3000 ng/ml, false negative results were observed with AbuSign(TM) kit. While different concentrations of alum in urine have no effect on TDxR immunoasays, nevertheless it exhibited significant infulence on AbuSign(TM) and EmitR assays as well. |