| 英文摘要 |
Timely and accurate laboratory reports are essential for diagnosis in emergency care; however, in large hospitals, high patient volumes often prolong patient waiting times. This study aimed to enhance the efficiency of hematology testing by optimizing the staining procedure for blood cell morphology differentiation. The total staining time of the ''FANS'' Wright Giemsa stain was reduced from 11 minutes 33 seconds to 4 minutes 54 seconds, representing a 2.4-fold increase in efficiency. The staining quality met the standards for manual interpretation, and the accuracy of automated image-based cell recognition reached 85.6%, surpassing the 84% reported in previous literature. Analysis of urgent specimens collected between 2020 and 2024 showed no significant difference in sample volume; however, the mean monthly turnaround time decreased from 29.62 minutes to 25.56 minutes, reflecting a 13.7% improvement. Furthermore, the report revision rate decreased by 2.5-fold, and 75% of medical technologists expressed satisfaction with the staining results. These findings demonstrate that the optimized“FANS”Wright–Giemsa staining protocol effectively improves both efficiency and diagnostic quality, supporting its potential for clinical application. |