英文摘要 |
From September 2018 to April 2019, the infection control center conducted aerobic bacterial count and adenosine triphosphate bioluminescence (ATP) inspection for medical equipment. The average pass rate of microbial cultivation and detection by suction machine in this unit is 27.5%, and the average pass rate of ATP bioluminescence is 2.5%. Through a variety of improvement measures, including discussion with infection control nurses about the timing, frequency, and steps of cleaning sputum suction machine ; Planning 5S team personnel to benchmark units to observe and learn the sputum suction machine cleaning videos after learning to let unit staff know the cleaning steps ; Standardize the cleaning frequency of sputum suction machine into the unit´s 5S team and audit; hand hygiene education; audit the suction process to achieve the purpose. From May to November 2019, the average qualification rate of the microbial cultivation and the ATP bioluminescence was 90% and 80%. |