Purposes: General radiologic images are an essential technique in diagnostic radiology. Image quality and the relative dose of general radiologic images are thus areas of interest. Methods: This study used Taguchi method L9(34) to optimize the operative parameters in abdominal X-rays to achieve better images from a practical viewpoint. An acrylic dual phantom was adopted in this work because the pellet prosthesis and hollow cylindrical phantom had similar responses to air and tissues in the actual examination and a higher SNR; the conditions changed and affected the results because the hospital had set the conditions. Results: Based on ANOVA, the experimental results showed that the optimal parameter settings were as follows: kVp, 70; mAs, 30; FOV, 14 inches x 17 inches; and SID, 80. The reconstruction filter for smooth and ring artifact suppression was weak and verified based on ANOVA. Conclusions: The optimal setting had a superior SNR or a lower dose to the original setting. Using Taguchi method L9(34) not only minimized the experimental frequencies, but also lowered the exposed dose in the actual examination. |