Ethylene Oxide (EtO) is widely used to sterilize heat and moisture-sensitive materials in hospitals. It is also used as a fumigant in the textile and food industries. EtO is found to be a carcinogen, therefore, its permissible exposure level is lowered in recent years. It becomes necessary to develop new sampling and analytical methods of measuring EtO at low concentrations in order to estimate potential health effects from EtO in the occupational environment. This study is designed to develop a sampling and analytical method for measuring EtO exposures at ppm levels. We have successfully demonstrated that a sampling Carbosieve III sampling tube (600 mg) and a low-flow pump (60 mL/min for 6 hours) is able to collect enough amounts of EtO. The adsorbed EtO can be thermally desorbed by a thermal desorption unit and analyzed by a GC/FID system (HP5890). In analysis, EtO was first thermally desorbed atl00°C (Tekmar Inc., Model 6000) from Carbosieve and then cryogenically focused by liquid nitrogen at -150°C and finally carried into a 60m long Supelcowax 10 capillary column for separation and a flame ionization detector for identification and quantification. For EtO, the desorption efficiency of thermal desorber was 100%, the detection limit was 1.2μg, and the storage stability was 9 days at -10°C in the refrigerator. |