Ondansetron is a novel 5-hydroxy-tryptamine3 (5-HT3)receptor antagonist which is a drug used clinically for chemotherapy-, radiotherapy-and postoperation induced nausea and emesis. A simple, sensitive, precise and accurate HPLC method for the detemination of ondansetron and its metabolites 7-hydroxy-ondansetron and 8-hydroxy-ondansetron in plasma is reported. Samples were prepared by a traditional, inexpensive solvent extraction method. The method employed a CN analytical column with an isocratic mobile phase of 20% acetonitrile in 80% 0.0125 M NH4OAc (pH = 4.7) and ultraviolet detection at 305 nm. Using this method, the calibration curves of ondansetron, 7-hydroxy-ondansetron and 8-hydroxy-ondansetron in plasma in the range of interest (20-800 ng/ml, 100-800 ng/ml and 100-1000 ng/ml, respectively) showed good linearity (r >0.999) and precision(both between-day and within-day coefficients of variation less or around 10%). The lower detection limits of this assay method for ondansetron and its 7-OH and 8-OH metabolites were 5 ng/ml, 10 ng/ml and 10 ng/ml respectively. This method has been successfully applied to the pharmacokinetic study of ondansetron in cancer patients receiving single i. v. infusion followed by a multiple oral dose of ondansetron. |