篇名 | 降低血液透析逆滲透水處理系統與透析液菌落數之專案 |
---|---|
並列篇名 | A Project for Reducing Reverse Osmosis Water Conditioning Systems and Bacteria Count in Hemodialysate |
作者 | 余秀芬、張麗娟 |
中文摘要 | 透析醫療品質的首要重心是透析液,本單位於2010 年4 月至12 月透析液菌落數偏高,確保病人安全故成立專案小組,旨在降低逆滲透(reverse osmosis, RO)水處理系統與透析液菌落數、提升透析護理品質。經分析發現問題有:透析機消毒時間不足、RO 水處理系統採水時取樣口污染、透析機RO 水入口處未執行無菌技術採水、人員對透析液品質知識不足、透析機老舊。改善方法為:(一)修訂透析室採水標準技術;(二)修訂水質異常處理作業規範、制訂設備管理作業規範、制訂RO 設備消毒作業規範;(三)更換老舊血液透析機;(四)建立稽核機制;(五)舉辦在職教育等方法,結果菌落數異常件數3.92 件/月降至0 件/月,顯示病人安全得到保障也提昇了透析品質。 |
英文摘要 | Dialysate is regarded as the key in the healthcare quality of dialysis. The elevated dialysate bacteria count of this unit between April and December of 2010 led to the establishment of an ad hoc group assigned to reducing the bacteria count in the reverse osmosis (RO) water conditioning systems and dialysate, as well as enhancing the quality of nursing care during dialysis. Some of the problems discovered through the analysis included inadequate sterilization of dialysis machines, sampling port pollution of RO water conditioning systems, failure to use aseptic water intake techniques for the RO water inlets of dialysis machines, insufficient knowledge about dialysate quality, and worn dialysis machines. Suggested improvements included: (1) revising the standard technique for water intake in dialysis units; (2) revising the standard operating procedures for handling abnormal water quality, equipment management, and RO apparatus disinfection; (3) replacement of worn hemodialysis machines; (4) establishing auditing mechanisms, and (5) hosting of inservice education. The number of cases of abnormal bacteria count subsequently dropped from 3.92 to 0 cases per month. Patient safety was therefore assured while dialysis quality was improved as well. |
起訖頁 | 011-025 |
關鍵詞 | 病人安全、透析液、菌落數、patient safety、dialysate、bacteria count |
刊名 | 臺灣腎臟護理學會雜誌 |
出版單位 | 臺灣腎臟護理學會 |
期數 | 201603 (15:1期) |
DOI | 10.3966/172674042016031501002 複製DOI DOI申請 |
QRCode | |