英文摘要 |
A gas chromatography-mass spectrometry method for the analysis of phencyclidine (PCP) in urine was subjected to an interlaboratory study. The collaborative study followed the guidelines provided by the Association of Official Analytical Chemists International. Ten laboratories participated, analyzing 3 samples of PCP-spiked urine as blind duplicate. The repeatability relative standard deviation (RSDr) and the reproducibility relative standard deviation (RSDR) were between 2.1%-3.6% and 4.2%-7.3%, respectively. HORRAT values for the reproducibility showed 0.4-0.7, indicating acceptable precision between laboratories. The method was thus proposed to be used by the drug-abuse urine testing laboratories in Taiwan. |