英文摘要 |
The inclined air-curtain (IAC) fume hood employed in the present work was constructed with the cross-flow fan at the bottom, deflection plate along the side walls, and a suction slot at the top of the cabinet. The containment in the cabinet was controlled effectively by using the two-dimension blow velocity and suction velocity. A research of an IAC walk-in fume hood performance was operated in several methods. Face velocity was measured at different sash opening to know the distribution and it’s average face velocity. The mean leakage SF6 concentration is less than 0.005 ppm at p1~p8 block by inner test method based on EN14175-3. The peak concentrations are 0.027 ppm. The tracer gas concentrations are less than 0.005 ppm in 10 cm,15 cm, 20 cm and 25 cm sash opening by self-develop method. The average concentration are less than 0.005 ppm, peak concentration are 0.011 ppm and 0.010 ppm at 30 cm and 35 cm sash opening. The average concentration of SF6 is 0.093 ppm, peak concentration is 0.280 ppm at Hej=60 cm, H=20 cm~100 cm sash movement test. The tracer gas test results in different blow jet velocity were as below: Vb=1.7 m/s, Cavg=0.056 ppm; Vb=0.7 m/s, Cavg= 0.115 ppm; Vb=0.2 m/s, Cavg=0.179 ppm. Robustness test results were: The average concentration is 0.041 ppm when Vb is 1.7 m/s. The average concentration is 0.155 ppm when Vb is 0.7 m/s. When the blow jet velocity (Vb) decreased, the containment efficiency also reduced at sash movement test and robustness test. This result showed that an adequate blow jet velocity would provide a higher containment performance. |